AMETEK SL Series Software User Manual
Page 29

Elgar Electronics Corporation
Calibration Procedure
1.
a) Set the DC Power supply to the value shown in the DC INPUT
column of Table 2-18;
b) set the electronic preset value to the value shown in the Setting
column of Table 2-18.
2.
PRES OFF (press PRES key until its LED is not lit), so that the DVM
is in Measurement mode and not Preset mode.
3.
LOAD ON (press LOAD key until its LED is lit); adjust using the
keys until DMM reading matches the value shown in DMM Reading
column of Table 2-18.
4.
Scroll to next calibration parameter (press
), Item 14, and repeat CV
Steps 1 through 3.
5.
Press
to return to Item 13, and repeat CV Steps 1 through 4 until
no further adjustments are required for both Items 13-14.
ITEM Setting DC
INPUT
DVM
13
0.200 V
61.00 V / 0.1A
0.2000 V
14
60.00 V
61.00 V / 0.1A
60.00 V
*
80.00 V
81.00 V / 0.1A
80.00 V
*
Applies to SLD-80-20-102 only
Table 2-18. Channel B CV Mode Calibration Settings
6.
Scroll to next calibration parameter, Item 15 (press
).
SLD Calibration Manual
2-21