Calibration Procedure

CR Range I Offset and Gain, Channel A CR Mode

1.

Set load to match the value in CR Range I Offset column of Table 2-8,
for the model being calibrated.

2.

Set the DC Voltage Standard to match the value in IV Sense Input
column of Table 2-8 for the same model being calibrated.

3.

LOAD ON (press LOAD key until its LED is lit).

4.

Adjust using

…† keys until DMM reading matches the value in DMM

Reading column of Table 2-8.

ITEM MODEL

CR Range I OFFSET V SENSE input

DMM Reading

SLD-60-505-255

4500

Ω

60.00 V

0.013 A

SLD-61-505-255

4500

Ω

60.00 V

0.013 A

SLD-60-20-102

11250

Ω

60.00 V

0.0053 A

SLD-80-20-102

15000

Ω

80.00V 0.0053

A

SLD-61-5-752

45000

Ω

60.00 V

0.0013 A

SLD-62-5-752

45000

Ω

60.00 V

0.0013 A

15

SLD-60-105-550

2250

Ω

60.00 V

0.026 A

Table 2-8. Channel A CR Mode Range I Offset Calibration Settings

5.

Scroll to next calibration parameter, Item 16 (press

„).

6.

Set the DC Voltage Standard to match the value in V Sense Input
column of Table 2-9 for the same model being calibrated.

7.

LOAD ON (press LOAD key until its LED is lit). Adjust using

…† keys

until DMM reading matches the value in DMM Reading column of
Table 2-9.

8.

Repeat Steps 1 through 7 (press

ƒ to return to previous item) until no

further adjustments are necessary for the DMM reading to match the
value in the DMM Reading column for both calibration parameters
(Items 15 and 16).

ITEM MODEL CR Range I GAIN V SENSE input

DMM Reading

SLD-60-505-255

1.200

Ω

15.00 V

12.50 A

SLD-61-505-255

1.200

Ω

15.00 V

12.50 A

SLD-60-20-102

3.000

Ω

15.00 V

5.000 A

SLD-80-20-102

4.000

Ω

15.00V 3.750

A

SLD-61-5-752

12.000

Ω

30.00 V

2.500 A

SLD-62-5-752

12.000

Ω

30.00 V

2.500 A

16

SLD-60-105-550

0.600

Ω

6.000 V

10.00 A

Table 2-9. Channel A CR Mode Range I Gain Calibration Setting

SLD Calibration Manual