Table 2

Guidance and manufacturer´s declaration – electromagnetic immunity
The BC40 Wrist Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of

the BC40 Wrist Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test

IEC 60601 test level

Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)

IEC 61000-4-2

± 6 kV contact

± 8 kV air

± 6 kV contact

± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are

covered with synthetic material, the relative humidity should be

at least 30 %.

Electrostatic transient / burst

IEC 61000-4-4

± 2 kV for power

supply lines

± 1 kV for input/output

lines

Not applicable

Mains power quality should be that of a typical commercial or

hospital environment.

Surge

IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

Not applicable

Mains power quality should be that of a typical commercial or

hospital environment.

Voltage dips, short interrup-

tions and voltage variations on

power supply input lines

IEC 61000-4-11

± 1 kV differential mode

± 2 kV common mode

Not applicable

Mains power quality should be that of a typical commercial or

hospital environment. If the user of the BC40 Wrist Digital Blood

Pressure Monitor requires continued operation during power

mains interruptions, it is recommended that the BC40 Wrist Digi-

tal Blood Pressure Monitor be powered from an uninterruptible

power supply or a battery.