Beurer BC 40 User Manual

Page 103

Advertising
background image

103

Table 3

Guidance and manufacturer´s declaration – electromagnetic immunity
The BC40 Wrist Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of

the BC40 Wrist Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test

IEC 60601 test

level

Compliance

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 V

rms

150 kHz to

80 MHz

3 V/m 80 MHz to

2,5 GHz

Not applicable

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part

of the BC40 Wrist Digital Blood Pressure Monitor, including cables, than the recommen-

ded separation distance calculated from the equation applicable to the frequency of the

transmitter.

Recommended separation distance:

80 MHz to 800 MHz

800 MHz to 2,5 GHz

where p is the maximum output power rating of the transmitter in watts (W) according to

the transmitter manufacturer and d is the recommended separation distance in metres

(m).b

Field strengths from fixed RF transmitters, as determined by an electromagnetic site

survey,a should be less than the compliance level in each frequency range.b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures,

objects and people.

P

V

d

]

5

.

3

[

1

=

P

V

d

]

5

.

3

[

1

=

P

E

d

]

7

[

1

=