Beurer BC 40 User Manual

Page 102

Advertising
background image

102

Guidance and manufacturer´s declaration – electromagnetic immunity
The BC40 Wrist Digital Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of

the BC40 Wrist Digital Blood Pressure Monitor should assure that it is used in such an environment.
Immunity test

IEC 60601 test level

Compliance level Electromagnetic environment – guidance

Power frequency (50/60 Hz)

magnetic field

IEC 61000-4-8

< 5 % UT

(>95 % dip in UT )

for 0.5 cycle

40 % UT

(60 % dip in UT )

for 5 cycles

70 % UT

(30 % dip in UT )

for 25 cycles

< 5 % UT

(>95 % dip in UT )

for 5 sec

3 A/m

Power frequency magnetic fields should be at levels charac-

teristic of a typical location in a typical commercial or hospital

environment.

NOTE 1 U

T

is the a. c. mains voltage prior to application of the test level.