• you move or speak during the measurement ( appears

on the display)

• the cuff is placed on the arm too tightly or loosely or if the

cuff is damaged ( appears on the display)

• the pump pressure is higher than 290 mmHg (

appears on the display)

• the batteries are almost worn out ( and

).

In such cases, repeat the measurement. If necessary, rein-

sert or replace the batteries.

8. Cleaning and maintenance

• Clean your blood pressure monitor carefully using a

slightly damp cloth only.

• Do not use any cleaning agents or solvents.

• Under no circumstances should you hold the unit under

water, as this can cause liquid to enter and damage the unit.

• Never place any heavy objects on the device.

9. Technical specifications

Model no.

BC 40

Measurement

method

Oscillometric, non-invasive blood pressure

measurement on the wrist

Measurement

range

Cuff pressure 0 – 290 mmHg,

Systolic 50 – 250 mmHg,

Diastolic 40 –180 mmHg,

Pulse 40 – 180 beats/minute

Display accu-

racy

Systolic ± 3 mmHg, diastolic ± 3 mmHg, pulse

± 5 % of the value shown

Measurement

inaccuracy

Max. permissible standard deviation accord-

ing to clinical testing: systolic 8 mmHg/dias-

tolic 8 mmHg

Memory

60 memory spaces

Dimensions

L 76 mm x W 70 mm x H 23 mm

Weight

Approximately 105 g (without batteries)

Cuff size

12.5 to 21.5 cm

Permissible

operating con-

ditions

+ 10 °C to + 40 °C, ≤ 85 % relative humidity

(non-condensing)

Permissible

storage condi-

tions

- 20 °C to + 55 °C, ≤ 85 % relative humidity,

800-1060 hPa ambient pressure

Power supply

2 x 1.5V

AAA batteries

Battery life

For approx. 300 measurements, depending on

levels of blood pressure and pump pressure

Accessories

Instructions for use, 2 x 1.5V AAA batteries,

storage box

Classification

Internal supply, continuous operation, applica-

tion part type BF, IPX0, no AP or APG