21

• the inflation pressure is higher than 300 mmHg (ERR 300),

• the batteries are almost empty

.

In such cases, repeat the measurement. Ensure you do not

move or speak. If necessary, re-insert or replace the batteries.

9. Specifications

Model No.

BC 60

Method of

Oscillating, non-invasive blood pressure

Measurement

measurement on the wrist

Range of

Pressure 0 – 300 mmHg,

Measurement

pulse 40 – 160 beats/minute

Accuracy of the

systolic ± 3 mmHg / diastolic ± 3 mmHg /

pressure reading

pulse ± 5 % of the displayed value

Measuring

max. permissible standard deviation

uncertainty

according to clinical testing:

systolic: 8 mmHg / diastolic: 8 mmHg

Sets of Memory

2 x 60 memory places

Unit Dimension

L 72 x W 80 x H 28.7 mm

Unit Weight

Approx. 140 g

Cuff Size

For a wrist circumference of approx.

125 to 215 mm

Operating

Temperature

+10 °C to +40 °C, ≤ 85 % RH

Storage-Temperature -5 °C to +50 °C, ≤ 85 % RH

Power Supply

2 x 1,5 V Micro (Alkaline Type LR 03)

Battery Life

For 250 measurements

Accessories

Storage box, instructions for use,

2 x AAA batteries

Classification

Application part Type BF

Key to symbols

Application part Type BF

Important! Read the operating

instruction!

These specification are subject to change without notice for

purpose of improvement.
• This unit is in line with European Standard EN60601-1-2

and is subject to particular precautions with regard to

electromagnetic compatibility (EMC). Please note that

portable and mobile HF communication systems may in-

terfere with this instrument. For more details, please con-

tact customer service at the address indicated.

• The instrument is in line with the EU Medical Products

Directive 93/42/EC, the German medical products act

and European standards EN1060-1 (non-invasive blood

pressure monitors, Part 1: general requirements) and

EN1060-3 (non-invasive blood pressure monitors, Part

3: Supplementary Requirements for Electromechanical

Blood Pressure Measurement Systems).

• If you are using the instrument for commercial or eco-

nomic purposes, you must check measurement precision

regularly in accordance with the ’Operators’ Ordinance

on Medical Products’. Even in the case of private use, we

recommend checking measurement precision every two

years at the manufacturers.