Commissioning Procedure

Check wiring
1. Make sure that the conductivity transmitter system is wired in accordance with the

wiring diagram.

2. Check that the supply voltage corresponds to the name plate rating.

Set code switch
The transmitter is parameterised by means of a ten-pole code switch.
This code switch can also be used to adjust the cell constant and trigger off the functional
test.
The individual switch positions can be changed by using e. g. a ball pen.

Factory setting

S1 S2 S3 S4 S5 S6 S7 S8 S9 S10
ON ON ON OFF ON ON OFF OFF OFF OFF

In the following table the factory setting is highlighted in grey:

Adjust measuring length (control range) and actual value output

Code switch

Control range

Current output mA =

µS/cm

1 2 3 [

µS/cm at 25 °C]

4 mA correspond to

20 mA correspond to

OFF OFF OFF

20

20

ON

OFF OFF

100

100

OFF ON

OFF 200

200

ON ON OFF 500

500

OFF OFF ON

1000 1000

ON

OFF ON

2000 2000

OFF ON ON

6000 6000

ON

ON

ON

0.5

12000

0.5

12000

During the commissioning procedure (when the supply voltage is applied) the first output
will be 4 mA, then the output current increases until it reaches the actual value.

Check temperature coefficient Tk setting

For a linear compensation of the measured conductivity value based at 25 °C the
temperature coefficient Tk is set at our works to 2.1 % per °C. Once the service
temperature is reached you can check this setting by carrying out a comparison
measurement - e. g. as part of the commissioning procedure. If there is a deviation
between the indicated conductivity and the measured comparison value you can correct
the measurement result of the transmitter by adjusting a lower or higher temperature
coefficient. Continue modifying the Tk setting until the indicated conductivity value and
measured comparison value agree.

Code switch

Temperature coefficient
Tk [%/°C]

4 5 6 7

OFF OFF OFF OFF

0 (no compensation)

ON

OFF OFF OFF

1.6

OFF

ON

OFF OFF

1.7

ON ON OFF OFF

1.8

12