Reducing length of installation and measuring length (control range)
The measuring tube and the electrode can be cut to suit the individual requirements of the
installation.

Length of installation A

[mm]

Measuring length L

[mm]

Possible reduction

of length R [mm]

215 200

315 300 50
415 400 50
515 500 50
615 600 50
815 800 100

1015 1000 100

1. Carefully determine the required reduction of length - measured from the bottom end of

the measuring tube - and mark this on the measuring tube.

2. Only slacken locking grub screw using a screwdriver size 1.3 for socket-head cap

screws.

3. Unscrew measuring tube and cut it to the length marked.

4. Unscrew spacer disc.

5. Mark the reduction of length - measured from the extreme end of the tip - on the

electrode tip.

6. Cut electrode tip to length. Cut back 10 - 12 mm of the insulation.

7. Screw back the spacer disc onto the electrode tip and fasten fingertight with a 12 mm

open end spanner.

8. Screw back measuring tube with a torque of 50 Nm and tighten locking grub screw.

Attention

!

The seating surfaces and threads on the vessel and mounting flanges must be
accurately machined.

!

Do not bend the electrode tip when cutting. Do not subject electrode to physical
shocks.

!

Do not lag the electrode body.

!

The conductivity transmitter system can be installed horizontally or with a vertical
inclination. The measuring surface must be permanently submerged.

Tools
!

41 mm open-end spanner

!

12 mm open-end spanner

!

Screwdriver size 1.3 for socket-head cap screws

!

Hacksaw

!

Flat file, cut 2

6