Flowserve LRGT16-1 Spector compact Conductivity Transmitter System User Manual
Page 17
Connection
1" BSP to DIN ISO 228
Materials
Screwed
housing:
1.4571
Measuring tube/electrode/screw:
1.4571
Electrode rod insulation:
PTFE
Spacer
disc:
PEEK
Terminal
box: 3.2161
aluminium
Length of installation / Measuring length / Reduction of length
Length of installation A
[mm]
Measuring length L
[mm]
Possible reduction of
length R [mm]
215 200
315 300 50
415 400 50
515 500 50
615 600 50
815 800 100
1015 1000 100
Temperature sensor
Resistance thermometer Pt 1000
Cell constant C
0.2 cm
-1
Wiring
5pole screw-type terminal strip, removable, conductor size 1.5 mm² or via 5pole connector
Cable entry
Cable gland with integral cable clamp, 1 x PG 9 (M16), 1 x PG 16 (M20)
Protection
IP 65 to DIN 40050
Max. admissible ambient temperature
70 °C
Temperature for storage and transport
– 40 to + 80 °C
Supply voltage
24 V DC (18 - 36 V DC)
17