Wiring

Connect conductivity transmitter

The equipment has to be supplied with 24 V DC from a safety power supply unit (PSU) in
accordance with DIN VDE 0106 (safe supply isolation). The PSU must be equipped with
an overcurrent protective device according to DIN EN 61010-1/VDE 0411.

The following steps 1 - 5 do not apply for equipment fitted with a connector.

For wiring use flexible screened control cable, min. conductor size 0.75 mm².

1. Remove terminal strip and pull connecting cable through cable gland.

2. Strip off approx. 40 mm of cable insulation coating and remove approx. 5 mm of

conductor end insulation.

3. Carry out wiring in accordance with wiring diagram.

4. Plug in terminal strip.

5. Seal cable entry by tightening cable glands. Use sealing plug to seal off cable glands

that are not used.

6. Install terminal cover and tighten cover screws, ensuring correct seat of cover gasket.

Tools

!

Screwdriver size 1 for cross-recess head screws

!

Screwdriver size 2.5 for slotted screws, completely insulated to VDE 0680

!

18 (19) mm open-end spanner