Troubleshooting, Fault finding list, Equipment does not work – Flowserve LRGT16-1 Spector compact Conductivity Transmitter System User Manual

Page 15

Advertising
background image

Troubleshooting


Fault Finding List
Equipment does not work
Fault:

Supply voltage failure - no function.

Remedy:

Check power supply and all electrical connections.


Fault:

Electronic insert defective.

Remedy:

Replace electronic insert.

Equipment signals malfunction
LED display:

Red LED flashing.

Current output: 4 mA
Fault:

Value below 0 % setting, e. g. electrode exposed.

Remedy:

Check water level and installation.


LED display:

Red LED flashing.

Current output: 0 mA
Fault:

Connection to measuring tube/electrode interrupted.

Remedy:

Check electrode cables (electronic insert, terminal lugs 1 - 4).
Replace equipment if necessary.


LED display:

Green LED flashing.

Current output: 0 mA
Fault:

Value above 100 % setting, e. g. control range too small.

Remedy:

Adjust larger control range.


LED display:

Red LED flashing.

Current output: 0 mA
Fault:

Connections to measuring tube/electrode short-circuited.

Remedy:

Check electrode cables (electronic insert, terminal lugs 1 - 4).
Replace equipment if necessary.


LED display:

Red and green LEDs flash alternately.

Current output: 0 mA
Fault:

Temperature in terminal box exceeds 85 °C.

Remedy:

Check ambient temperature. Make sure it does not exceed 70 °C.


LED display:

Red and green LEDs flash alternately.

Current output: 0.5 mA
Fault:

Electrical connections to resistance thermometer interrupted or
short-circuited or thermometer defective.

Remedy:

Check thermometer connections (electronic insert, terminal lugs 5 + 6).
Replace equipment if necessary.