LED Display and Fault Indications
A green and a red LED for status and fault indication are provided in the centre of the
electronic insert.

Normal operation

green LED red LED

Current output

[mA]

Conductivity 0 to +10 %
of control range

lit

proportional

to

measured value

Conductivity 90 to 100 %
of control range

lit

proportional

to

measured value

Conductivity 10 to 90 %
of control range

lit lit

proportional

to

measured value

Fault

green LED red LED

Current output

[mA]

Value below 0 % setting,
e. g. electrode exposed

flashes

4

mA

Connections to measuring
tube/electrode interrupted

flashes 0

mA

Value above 100 % setting,
e. g. control range too small

flashes

0

mA

Connections to measuring
tube/electrode short-
circuited

flashes 0

mA

Temperature in terminal box
exceeds 85 °C

flashes flashes

0

mA

Connection to resistance
thermometer interrupted or
short-circuited

flash alternately

0.5 mA