Appendices, System setup checklist – Flowserve StarPac II Control Valves User Manual

Page 34

Advertising
background image

Appendices

System Setup Checklist

A. Power

• (24) VDC power at least 300 mA.
• Polarity

correct.

• Local display ON. If not, check power supply.

B. System Communications (if connected to a remote computer)

• RS-485 converter properly installed or attached to computer.
• Signal polarity correct.
• Modbus device address set in StarPac device; refer to StarPac Operation section.
• Start software; refer to software section.
• Configure communications in software and hardware to match. Refer to Initial Setup in The

StarTalk II portion of this manual and StarPac Operation section.

• Communications work. If not, recheck settings.

C. Calibration

checks

• Air supply turned on.
• Check status screens to verify system calibration; refer to software section.
• Valve stroke calibration to remove installation and handling offsets; refer calibration section.
• Verify process sensor calibration; recalibrate if needed.

D. System Configuration; refer to engineering/configuration section for details

• Set or verify failure modes.
• Set or verify analog output.
• Set or verify command and mode source
• Set or verify stroke limits.
• Set or verify LCD display options.
• Check positioner response and set gain to control speed.

E. Automatic PID Operation; refer to Engineering/Tuning section

• Select process variable.
• Input full scale range.
• Set initial PID parameters.
• Begin Automatic mode and tune system.

F. Other Options; refer to Engineering section

• Set

units.

• Reset

totalizer.

• Set or verify tagname.
• Set and start Data Logger function.
• Collect and save installed signature, if desired.

G. Save Installed Configuration; refer to technician support section

• Save

configuration.

• Make backup copy and archive.

StarPac II User Interface Manual

28

Advertising