Commissioning procedure – Flowserve OR 52-7 User Manual

Page 19

Advertising
background image

9

The following parameters can be selected and, if required, changed:
Value Offset (reduction of zero point 0-5 ppm),

Output (Output of actual value 0 / 4-20 mA),

Alarm  Limit Value (Limit Alarm , 0-5 ppm),

Alarm 2 Limit Value (Limit Alarm 2, 0-5 ppm),

Alarm  Time Delay (Delay of response for alarm , 0-5 sec),

Alarm 2 Time Delay (Delay of response for alarm 2, 0-5 sec),

Datalog Export (transferring the storage data to a PC by means of PC software).
. Press key

P to enter the programming mode. The amber

system LED will be illuminated and the window showing

the parameter that was changed last appears.

2. Press  or  to scroll forwards / backwards until the

desired parameter is shown.

3. Press key

P (P = change) to activate the parameter chan-

ge. To quit the programming mode without any changes

press key OK (OK = back). The window that was active

before you switched the mode will reappear.

4. After having activated the parameter change by pressing

key

P use the keys +/ to adjust the value within the

permissible limits.

5. To cancel the new setting press key

P (P = abort). The old

value will then remain valid.

6. To confirm and activate the new value press key OK

(OK = confirm). The parameter window appears again.

7. Press key OK (OK = back) to quit the programming mode.

The window that was active before you switched the mode

will reappear.

Programming mode

Commissioning Procedure

– continued –

Parameter
act Value Offset
Value : 0 ppm

= prev

= next

= back

OK

P = change

Parameter
act Value Offset
Value : 2 ppm

= inc Para

= dec Para

= confirm

OK

P

+

= abort

Parameter
act Value Offset
Value : 2 ppm

= prev

= next

= back

OK

P = change