Push cleaning device

4 Fig. 5 slowly into the mid position in order to simulate a malfunction.

The amber system LED is flashing and the relay contact Alarm 2 opens. Error code e or e2 is

indicated.
Push cleaning device

4 Fig. 5 back into its initial position.

The equipment must switch back to normal operation.

Functional test

Start, Operation, Alarm and Function Test

– continued –

If the alarm limits  and 2 are exceeded, the adjusted time delay is activated.
LED Alarm  and LED Alarm 2 are flashing. After the time delay LED Alarm  and LED Alarm 2 are

permanently illuminated. The relay contacts Alarm  and Alarm 2 open.
Alarm 2 is also indicated when

the separator is switched off,

the purging valve Clw is opened and

if the sensor ORG 2 signals a malfunction

Depending on the reason that caused Alarm 2, the equipment

state “ppm>Limit”, “Separator OFF” or “Clear Water On” will

be indicated in the measurement window.

Alarm 1 and 2

OIL CONTENT :

ppm

State: ppm > Limit

16

UTC 10:46

2006-12-06