Start, operation, alarm and function test – Flowserve OR 52-7 User Manual
Page 22
22
Push cleaning device
4 Fig. 5 slowly into the mid position in order to simulate a malfunction.
The amber system LED is flashing and the relay contact Alarm 2 opens. Error code e or e2 is
indicated.
Push cleaning device
4 Fig. 5 back into its initial position.
The equipment must switch back to normal operation.
Functional test
Start, Operation, Alarm and Function Test
– continued –
If the alarm limits and 2 are exceeded, the adjusted time delay is activated.
LED Alarm and LED Alarm 2 are flashing. After the time delay LED Alarm and LED Alarm 2 are
permanently illuminated. The relay contacts Alarm and Alarm 2 open.
Alarm 2 is also indicated when
■
the separator is switched off,
■
the purging valve Clw is opened and
■
if the sensor ORG 2 signals a malfunction
Depending on the reason that caused Alarm 2, the equipment
state “ppm>Limit”, “Separator OFF” or “Clear Water On” will
be indicated in the measurement window.
Alarm 1 and 2
OIL CONTENT :
ppm
State: ppm > Limit
16
UTC 10:46
2006-12-06