Start, Operation, Alarm and Function Test

Start and operation

. Apply the supply voltage to the measuring transducer.

The green LED will be illuminated and alarm 2 is signaled.

2. When the equipment is switched on, the system window

appears for approx. 3 seconds, indicating the following:

Name of the equipment,

Certificate,

Serial number,

Number of the software version.

3. After approx 3 seconds the measurement window

appears, indicating the following:

Oil content (in large characters),

State: Warm up

Current time / date in UTC

4. After approx. 30 seconds the actual value in ppm is

indicated in the measurement window. If this value is

below 5 ppm the message Alarm 2 is cancelled and the

equipment state “Ready” is indicated. If, however, the

value continues to exceed 5 ppm, the message Alarm 2

remains active and the equipment state “ppm>Limit” is

indicated.

5. Press now key

+. The purging valve opens and Alarm 2 is

indicated. The equipment state “Clear Water ON” appears.

Rinse the measuring transducer ORG 2 a few minutes

with oil-free water and operate the cleaning device

4

2 – 3 times,

Fig. 5. During the rinsing process with oil-free

water the measuring window should show 0 ppm. If this

is not the case switch to programming mode, Parameter

Value Offset, and reduce the zero point by a max. of

5 ppm.

6. To close the purging valve press key

. The equipment

state “Ready” or “ppm>Limit” will be indicated.

Device
Certif.

Ser.Nr.

Vers.

: ORT7
: IMO

: MEPC.107(49)
: 392358

: 311137.10

Flowserve Gestra

OIL CONTENT :

ppm

State: WarmUp

8

UTC 10:46

2006-12-06

OIL CONTENT :

ppm

State: ready

8

UTC 10:46

2006-12-06

OIL CONTENT :

ppm

State: ClearWater ON

0

UTC 10:46

2006-12-06

Advertising