In the event of a malfunction the amber system LED is

flashing and the relay contact Alarm 2 is energised.

An error code is indicated in the measurement window.

OIL CONTENT :

ppm

State: e1 Light Ctrl ov

8

UTC 10:46

2006-12-06

The equipment is not working – no function, no indication

Fault:

The fuse of the equipment is defective.

Remedy: Replace the fuse of the equipment and check the power supply.

Fault:

Internal voltage supply failed.

Not all three green LEDs on the basis circuit board are illuminated.

Remedy: Check power supply or replace the measuring transducer.

Measured value increases continuously with 1ppm/day or slower

(ingress of foreign matter can be ruled out)

Fault:

Accumulated dirt deposits on glass cylinder.

Remedy: Operate the cleaning device more often.

Table listing error codes

Error code

Possibly caused by:

Remedy

e1 Light Ctrl ov Glow lamp in light emitter contaminated

Replace light emitter

Dirt deposits on glass cylinder

Clean glass cylinder

Excessive turbidity

Provide mechanical cleaning device upstream of

equipment

Operating voltage too low

Check voltage supply

Cleaning device not in initial position

Return the cleaning device into the initial position

Ingress of solid particles in fluid

Provide mechanical cleaning device upstream of

equipment

Error code

Possibly caused by:

Remedy

e2 no Light rec Glow lamp in light emitter defective

Replace light emitter

Connecting line to light emitter interrupted Check connecting line
Cleaning device not it its initial position

Return the cleaning device to its initial device

Ingress of solid particles in fluid

Provide mechanical cleaning device upstream of the

equipment

Error code

Possibly caused by:

Remedy

mem Card err

No or wrong memory card

Insert the right memory card

Once the fault is eliminated the equipment will return to normal operation.