Symbols, Manual objectives, Summary of features – Fluke Biomedical DPM1B User Manual

Page 13

Advertising
background image

Pneumatic Transducer Tester

Introduction

3

Symbols

Table 1 is a list of symbols found in this manual or on the Tester.

Table 1. Symbols

Symbol Description

Risk of Danger. Important information. See Manual.

Conforms to European Union directives

Conforms to relevant Australian EMC standards

Battery

This product complies with the WEEE Directive (2002/96/EC)
marking requirements. The affixed label indicates that you
must not discard this electrical/electronic product in domestic
household waste. Product Category: With reference to the
equipment types in the WEEE Directive Annex I, this product is
classed as category 9 "Monitoring and Control Instrumentation"
product. Do not dispose of this product as unsorted municipal
waste. Go to Fluke’s website for recycling information.

Manual Objectives

Information about the design of the Tester to enable the user to
understand its use.

Detailed guidance for maintaining and storing the unit.

Procedures for setting up and operating the Tester.

Summary of Features

The Tester is a hand-held pressure measuring device, designed as an
aid in the performance testing and troubleshooting of pneumatic or
hydraulic systems. A liquid crystal display (LCD) reading indicates
pressures manually generated from an internal chamber.

Testing blood pressure transducers is easy: Simply connect a BP
transducer to the luer lock fitting on the Tester and create any series of
pressures between

±

300 mmHg. Pressures can then be compared

directly between the Tester and the manometer or monitors under test.