Transducer sensitivity test – Fluke Biomedical DPM1B User Manual

Page 22

Advertising
background image

DPM1B
Operators Manual

12

Transducer Sensitivity Test

Only one measurement is required to determine whether or not a
transducer under test meets the manufacturer’s specified sensitivity.
The Tester is used to generate a known pressure (generally 100 mmHg
or higher works best). The excitation voltage of the transducer is
determined from the manufacturer’s specifications. Transducer
sensitivity is calculated as follows:

1. Apply 100 mmHg to the transducer as indicated by the Tester

LCD reading.

2. Record the pressure registered on the pressure monitor

3. Calculate the minimum and maximum pressure readings by

using the transducer’s sensitivity tolerance specifications.

For example: For a 100 mmHg applied pressure with a 1 %
tolerance, the minimum and maximum readings are 99 mmHg
and 101 mmHg, respectively.

4. Compare the readings from Step 2, along with the results from

Step 3. If the readings from Step 2 are within the range
calculated in Step 3, the transducer is within the manufacturer’s
specifications.

Linear Plot

Nonlinear Plot

200

150

DPM-1B
Readings
(mmHg)

Monitor Readings(mmHg)

100

50

50

100

150

200

ehf5.eps

Figure 5. Sample Plot of Transducer Linearity Using the DPM1B

Advertising