Testing transducer linearity – Fluke Biomedical DPM1B User Manual

Page 21

Advertising
background image

Pneumatic Transducer Tester

Testing Transducer Linearity

11

To interpret the results, proceed to the following three sections:

Testing Transducer Linearity

Transducer Sensitivity Test

Testing Other Devices

DPM-1B

Filter

2

1

Test Transducer

Vent Port

ehf4.eps

Figure 4. DPM1B Configured to Apply Pressure to Transducer

Testing Transducer Linearity

Several measurements are required to determine if a transducer is

linear over a range of pressure. See Figure 5 for a sample plot of

transducer linearity determined by using the Tester Pneumatic

Transducer Tester.

1. Test the transducer for linear operation in the range of

20 mmHg to 190 mmHg, for example, by taking measurements

in increments of 10 mmHg from 20 mmHg to 190 mmHg (as

shown in Figure 5).

2. Plot the resulting pressures by placing the Tester readings in

one axis and the monitor readings in the other axis.

3. Draw a line through the plotted points. A straight line indicates

that the transducer operates linearly; a curved line indicates

that the transducer operates nonlinearly.

4. Compare the actual pressure measurements with the

transducer manufacturer’s specifications for linear operation.

Advertising