Pneumatic Transducer Tester

Operation

Overview

9

Pneumatic

Transducer

Tester

BP Input

Filter

Monitor

2

1

ehf2.eps

Figure 2. DPM1B to Monitor Setup

3. When the system is to be filled with liquid, position Stopcock 1

as shown in Figure 2.

4. Attach a liquid-filled syringe to the top port of Stopcock 2.

5. Position the valve on Stopcock 2 so that the syringe can be used

to fill the system with liquids. (Turn the valve on the stopcock so
that the two opposing end arrows point straight up and down and
the middle arrow points toward the Tester, as depicted in
Figure 2.)

6. After filling the system, position both Stopcocks as shown in

Figure 2.

7. Remove syringe from top part of Stopcock 2.

Proceed to the section labeled Zeroing the Monitor and Tester.

Zeroing the Monitor and Tester

Properly zeroing the measurement system is important to accurately
determine transducer sensitivity.

1. Position the valve on Stopcock 1 as shown in Figure 4.

2. Press

the

ZERO or BALANCE control on the monitor.

3. Turn the Tester ON using the ON/OFF switch on the front

panel.