Operators Manual

6

Electromagnetic Compatibility (EMC)

International ........................................................ IEC 61326-1: Portable

CISPR 11: Group 1, Class A

Group 1: Equipment has intentionally generated and/or uses conductively-coupled

radio frequency energy that is necessary for the internal function of the equipment

itself.
Class A: Equipment is suitable for use in all establishments other than domestic

and those directly connected to a low-voltage power supply network that supplies

buildings used for domestic purposes. There may be potential difficulties in

ensuring electromagnetic compatibility in other environments due to conducted and

radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can occur when the

equipment is connected to a test object.

USA (FCC) .......................................................... 47 CFR 15 subpart B

Tester Familiarization

The Tester incorporates a highly accurate digital pressure meter and
pneumatic pressure-generating cylinder into a small, hand-held
package for determining the accuracy of most blood pressure
measurement systems.

Front Panel

The Tester incorporates a specially designed pressure cylinder and
precise solid-state transducer for generating and measuring static
pressure from

300 mmHg to

+

300 mmHg. The following items are

identified in Figure 1:

ON/OFF Switch: Used to power the Tester.

Thumbwheel: Used to regulate the pressure and pressure

value displayed on the LCD.

Luer Lock: Built-in to provide the external connection for

stopcocks or pressure fittings.

ZERO Knob: Adjusts LCD reading to zero.

Liquid Crystal Display (LCD): Illuminated when the Tester is -

powered. Displays LO BAT when the battery is low.

Note

When the Tester is first powered ON, the display may take a
few seconds to settle to zero.