Main switch on and off, Analyzer menu and function keys – Fluke Biomedical QA-ES II User Manual

Page 20

Advertising
background image

QA-ES II
Users Manual

10

Table 2. Controls and Connectors (cont.)

Item Name

Description

L

Terminals
RED and BLACK

Connection for the electrode outputs of the VAR. LOAD ESU. Attach the active electrode
to the red terminal and neutral electrode to the black terminal.

M

Fuses

T 200 mA at 230 VAC / T 400 mA at 115 VAC

N

Printer Port

25 pin D-sub

O

RS232 Serial Port

9-pin D-sub

P

Voltage Selector

115 VAC/230 VAC

Q

Mains Connector

3-pin connector to power cord

Main Switch On and Off

W

Caution

To avoid damaging the Analyzer, ensure the
rear-panel voltage selector switch is set to
the correct mains voltage and the proper
mains fuses are installed for the selected
voltage. Refer to items 16 and 13 in Figure 2
to check the voltage and fuses respectively.

The main switch is located on the front panel (see item 5
in Figure 1). When turning the Analyzer off, wait at least
five seconds before turning it on again to allow the reset
circuit to unload.

Analyzer Menu and Function Keys

The Analyzer provides flexibility and control over the
operations with a display, programmable function keys,
and a setting regulator. The upper part of the LCD screen
displays messages, status, and results. The menu bar is
at the bottom of the display. The function keys range from
Fl to F5. Select a function by pressing the key located
directly under the menu bar item.