Electrosurgical Analyzer

ESU Tests with the Analyzer

21

Measuring Leakage Current with Grounded HF
Equipment

With the ESU grounded, the test load is 200 ohms, and
the ESU must be operating at maximum power. Test
Setup C as shown in Figure 8 complies with IEC 601.2.2,
sec. 19.101a, test 1, fig. 102, and sec. 19.102, adopted
by ANSI/AAMI HF18-2001.

White

Fixed
Load

Measuring
Load

Black

Earthing

Contact

Red

White

Neutral

Active

QA-ES

ESU

eju013.eps

Figure 8. Test Setup C

Test Setup D shown in Figure 9 complies with IEC
601.2.2, sec. 19.101a, test 2, fig. 103, and sec. 19.102,
adopted by ANSI/AAMI HF18-2001.

White

Fixed
Load

Measuring
Load

Black

Earthing

Contact

Red

White

Neutral

Active

QA-ES

ESU

eju014.eps

Figure 9. Test Setup D

REM Alarm Test

The REM (Return Electrode Monitoring) alarm test
ensures that the ESU sounds an alarm if the resistance
between the two neutral electrodes exceeds its specified