Show choices (f1), Knob param. (f2), Start (f3) – Fluke Biomedical QA-ES II User Manual

Page 22

Advertising
background image

QA-ES II
Users Manual

12

SHOW CHOICES (F1)

This function is activated when you see an asterisk (*) in
the status field under ‘Mode’. Choose a test function by
pressing UP (F2) or DOWN (F3). You can also use the
encoder to choose a test function. Press ENTER (F5) to
save the test function under Mode in the STATUS field.
Press CANCEL (F4) to undo the selection.

F1

F2

F3

F4

F5

eju005.eps

KNOB PARAM. (F2)

With this function, you can choose among ‘Mode’, ‘Load’,
and ‘Delay’ in the STATUS field. An asterisk ‘*’ marks the
active item. If you choose ‘Load’, use the encoder to set
the load from 10 ohms to 5200 ohms in the following
increments:

In increments of 25 ohms from 50 ohms to 2500
ohms.

In increments of 100 ohms from 2500 ohms to 5200
ohms.

Save the selected load in ‘Mode’ under the ‘STATUS’
field by pressing ENTER (F5). Press CANCEL (F4) to
undo the selection.

If you choose ‘Delay’, use the setting regulator to set the
delay from 200 ms to 4000 ms in the following
increments:

In 50 ms increments from 200 ms to 1000 ms

In 100 ms increments from 1000 ms to 4000 ms

Save the chosen delay in ‘Delay’ under the ‘STATUS’
field by pressing ENTER (F5). Press CANCEL (F4) to
undo a choice.

START (F3)

Press START (F3) to start a test procedure. The text in
the field ‘Oper.’ changes from ‘Ready’ to ‘Measuring’. If
you have set the Analyzer to the position for a REM test,
this text changes from ‘Ready’ to ‘lncr. res.’ Press

STOP (F3) to stop the test procedure.

Advertising