Monopolar Testing

Cut Activation

1.

Connect the red jack on the Analyzer to the active
output on the ESU front panel.

2.

Select a proper load for cut output on the ESU.

3.

Connect the black jack on the Analyzer to the ESU
front panel via a dispersive return cable.

4.

Connect one of the Analyzer green jacks to the cut
activate pin of the ESU foot switch.

5.

Connect the other Analyzer green jack to the
common line of the ESU foot switch.

Coag Activation

1.

Connect the Analyzer red jack to the active output on
the ESU front panel.

2.

Select a proper load for the cut output on the ESU.

3.

Connect the Analyzer black jack to the ESU front
panel via a dispersive return cable.

4.

Connect one of the green jacks of the Analyzer to the
coag activate pin of the ESU foot switch

5.

Connect the other green jack on the Analyzer to the
common line on the ESU foot switch.

Bipolar Testing

1.

Connect the Analyzer red jack to the

load for the

bipolar cut

active output on the ESU front panel.

2.

Select a proper load for the

bipolar cut active

output

on the ESU.

3.

Connect the Analyzer black jack to the return of the
bipolar.

4.

Leave the dispersive cable connected to the ESU so
that REM does not alarm. Disconnect this single-pin
banana plug from the Analyzer and place it on a non-
conductive surface where it cannot contact anyone.

5.

Connect one of the Analyzer green jacks to the
activate pin of the ESU bipolar foot switch. Connect
the other Analyzer green jack to the common line on
the ESU bipolar foot switch.

Cleaning the Analyzer

Periodically wipe the Analyzer case with a damp cloth
moistened with a mild detergent. Do not use abrasives or
solvents.