Electrosurgical Analyzer

ESU Tests with the Analyzer

17

7.

After taking a measurement, the Analyzer resets
automatically and is then ready for additional testing.

If connected to a printer, the Analyzer writes a line of test
results to the printer following each Single Operation test.

Power Distribution Test

The Power Distribution test allows you to check the power
output performance of the ESU over a range of load
resistances. When using the Analyzer’s footswitch control
during the power distribution test, the ESU output power
automatically powers on and off. The ESU activates, the
Analyzer gets the reading on the first selected load, and
then the ESU turns off while the Analyzer changes the
load to the second selection. Next, the ESU powers on
again and the Analyzer gets the reading. This cycle
continues throughout the entire range of selected test
loads.

Connect the Analyzer and the ESU as shown in Figure 5
before starting the next series of tests.

2

1

Foot Switch

Green

Green

Black

Red

Neutral (dispersive)

Active

ESU

QA-ES

eju018.eps

Figure 5. Power Distribution Test