Foot switch control – Fluke Biomedical QA-ES II User Manual

Page 33

Advertising
background image

Electrosurgical Analyzer

Foot Switch Control

23

Foot Switch Control

There are three sets of two jacks each on the right side of
the Analyzer. The sets are as follows:

1.

FOOT SWITCH (green jacks)

2.

FIXED LOAD (white jacks)

3.

VAR. LOAD, (red and black jacks)

Note

For the following instructions, refer to the ESU
manufacturer’s instructions to determine the
wiring of the ESU monopolar and bipolar foot
switch jacks. Only qualified technicians should
do this, since the technician will be responsible
for making appropriate and safe test cables.

Refer to Figure 5 and Table 5 for connection examples.

Table 5. Foot Switch Connections

Monopolar Cut

A

Cut activate pin of ESU foot switch

B

Common

Coag

A

Coag activate pin of ESU foot switch

B

Common

Bipolar

A

Bipolar activate pin of ESU foot switch

B

Common