The Active electrode Test Setup A shown in Figure 6
complies with IEC 601.2.2, sec. 19.101b, fig, 104 and
sec. 19.102, adopted by ANSI/AAMI HF18-2001.

eju015.eps

Figure 6. Test Setup A

The Neutral electrode Test Setup B as shown in Figure 7
complies with IEC 601.2.2, sec. 19.101b, fig, 104 and
sec. 19.102, adopted by ANSI/AAMI HF18-2001.

Measuring
Load

Black

Earthing

Contact

Red

Neutral

QA-ES

ESU

eju016.eps

Figure 7. Test Setup B