Continuous operation test, Single operation test – Fluke Biomedical QA-ES II User Manual

Page 26

Advertising
background image

QA-ES II
Users Manual

16

Table 3. Load Resistance

Load Resistance Range

Equipment

IEC ANSI/AAMI

Monopolar

100 to 2000

Ω

50 to 2000

Ω

Bipolar

10 to 1000

Ω

10 to 1000

Ω

Continuous Operation Test

In the Continuous Operation mode, the Analyzer
continues to take measurements once you press the
START key. End this test by pressing the STOP key. The
Analyzer acts much like a meter during the test, showing
increasing and decreasing values as received from the
ESU.

To run a test in Continuous Operation mode:

1.

Press F2 until an asterisk (*) symbol appears
following ‘Mode’.

2.

Rotate the encoder knob until the *Cont. Oper test
mode appears on the screen, and then press the
Enter key.

3.

Set the test load as required for the test.

4.

Press START (F3) to begin testing.

5.

Activate the ESU and verify that the Analyzer is
recording measurements.

6.

Press STOP (F3) to end the test.

Single Operation Test

In the Single Operation mode, the Analyzer makes a
single measurement of the ESU output after the set delay
time. When the measurement is complete, the test
automatically stops. Make connections between the
Analyzer and the ESU as shown in Figure 4.

To run a test in Single Operation:

1.

Press F2 until an asterisk (*) symbol appears
following ‘Mode’.

2.

Rotate the encoder knob until the *Singl. Oper test
mode appears on the screen, and then press the
Enter key.

3.

Set the test Load as required for the test.

4.

Set the Delay time as required for the test.

5.

Set and activate the ESU.

6.

Press START (F3) to take a measurement. After the
set delay, the Analyzer records the values and
displays them on the screen.