High pressure, O2 measurement cell, Controls – Fluke Biomedical VT305 User Manual

Page 21

Advertising
background image

Gas Flow Analyzer

Start Up

7

High Pressure

The high pressure port is used to measure pressure more
than 200 mbars. Figure 4 shows the high pressure port on
the Product.

High

Pressure

gyo009.eps

Figure 4. High Pressure Port

Note

For measurements to a maximum of 200 mbar,
Fluke Biomedical recommends you use the
differential pressure port. Accuracy is 100 times
higher.

Caution

To prevent damage to the high-pressure
sensor, do not measure pressure more than
15 bar.

O

2

Measurement Cell

The Product has an interface for an O

2

measurement cell.

See Figure 5. For more information, see the O

2

Sensor

section in this manual.

O

2

– Measuring Cell

gyo010.eps

Figure 5. O

2

Cell

Controls

Table 3 is a list of the front-panel controls.