Configured and dchp ethernet setup, O2 sensor, Activation – Fluke Biomedical VT305 User Manual

Page 48: Installation, Oxygen sensor calibration – air only, Sensor

Advertising
background image

VT305
Users Manual

34

Configured and DCHP Ethernet Setup

The configured setup is used when a network that does
not have a DCHP server exists. The DCHP-Client setup is
used when you connect to networks that have a DCHP
server.

1. Use an Ethernet cable to connect the Ethernet port on

the Product to the network.

2. Push

X on the Product until the Ethernet Configured

or Ethernet DCHP – Client screen shows in the
Product display.

3. Open an Internet Browser.

4. In the address line, type the IP address shown in the

display of the Product and push Enter on the
computer keyboard.

Note

There can only be one connection to the
configuration tool for a single Product. With the
configuration tool open, the Product cannot be
configured from another computer.

The configuration tool will download to the computer and
establish a connection.

O

2

Sensor

Activation

The Product has an interface for an oxygen sensor. The
oxygen sensor must be calibrated with air and 100 % O

2

.

Installation

A kit that has the oxygen sensor and connection cable
comes with the oxygen option.

Remove the protective cap (rubber stopper) from the
sensor.

Oxygen Sensor Calibration – Air Only

Note

Fluke Biomedical does not recommend you
calibrate the oxygen sensor with air.

To calibrate the oxygen sensor with air:

1. Push

X on the front panel until O2 Calibration with Air

shows in the display.

2. Push

O to start the calibration process.

3. Apply 25 l/min of air to the flow channel of the Product

when the instruction shows in the display. See
Figure 31.

O2 Calibration

Air

Apply 25 I/min (0)

Press o to start

gyo066.eps

Figure 31. O

2

Calibration - Apply Air

4. Push

O to continue.