Setup for high pressure gases – Fluke Biomedical VT305 User Manual
Page 33
Advertising

Gas Flow Analyzer
Connect the Product
19
Setup for High Pressure Gases
The Product automatically compensates for the gas
pressure in the flow channel up to 150 mbar. Use the high
pressure port as shown in Figure 16 for pressures greater
than 150 mbar.
Caution
To prevent damage to the Product, do not
apply more than 800 mbars to the airway
channel port of the Product.
In the flow channel, the Product adjusts for pressures to a
maximum of 150 mbars. When the high pressure port is
used, the Product adjusts for pressures up to a maximum
of 300 mbars.
Flow Direction
Flow Direction
gyo051.eps
Figure 16. High Pressure Connections
Advertising