How to save data, Zero-point calibration – Fluke Biomedical VT305 User Manual

Page 28

Advertising
background image

VT305
Users Manual

14

The available filter selections are as follows:

None (Display of the latest measured value without

thresholds)

Low (Mean value over 240 ms)

Medium (Mean value over 480 ms)

High (Mean value over 960 ms)

The factory default for the filter is high.

You can change the filter selection in the browser-based
Product configuration tool. To learn more, see the Product
Configuration section.

How to Save Data

Push and hold O for 5 seconds to store data on the micro-
SD card. The screen in Figure 10 shows in the display
while the Product saves the data. See the How to Read
Out Measurement Data section.

Data saved to

DATAxx.CSV

gyo022.eps

Figure 10. Saving Data Screen

Zero-Point Calibration

Push and hold X for 5 seconds to start the zero calibration
of the pressure and flow sensors. While the Product does
the calibration procedure, the screen in Figure 11 shows in
the display.

Zero- Calibration

Running

gyo023.eps

Figure 11. Zero Calibration Screen

It is important to do a zero calibration periodically to
remove off-sets in the flow measurement.

Caution

To make accurate measurements, do not apply
pressure to the Product when you do a zero
calibration. This caution is not shown in the
display when you use the X symbol.

It is very important to do the zero calibration while the
airway pressure transducer stabilizes and before a
measurement is made.