Oxygen sensor calibration – o2 and air, Oxygen sensor calibration – o, And air – Fluke Biomedical VT305 User Manual

Page 49

Advertising
background image

Gas Flow Analyzer

O2 Sensor

35

Note

To stop the calibration procedure, push X.
Air calibration starts and will take 114 seconds to
complete. Under no circumstances interrupt the flow of
air through the flow channel. The screen in Figure 32
shows in the display when the calibration is done.

O2 Calibration

Air successful

Press any key to exit

gyo067.eps

Figure 32. O

2

Calibration Successful Screen

Oxygen Sensor Calibration – O

2

and Air

To calibrate the oxygen sensor with air and oxygen:

1. Push

X on the front panel until O2 Calibration with O2

and Air shows on the display.

2. Push

O to start the calibration process.

3. Apply 25 l/min of 100 % oxygen to the flow channel of

the Product when the instruction shows in the display.
See Figure 33.

O2 Calibration

100% O2

Apply 25 I/min (0)

Press o to contiue

gyo070.eps

Figure 33. O

2

Calibration - Apply Oxygen

4. Push

O to continue.

Note

To stop the calibration procedure, push X.
Oxygen calibration starts and will take 114 seconds to
complete. Under no circumstances interrupt the flow of
gas through the flow channel.

5. Apply 25 l/min of air to the flow channel of the Product

when the instruction shows in the display. See
Figure 31.

O2 Calibration

Air

Apply 25 I/min (0)

Press o to start

gyo066.eps

Figure 34. O

2

Calibration - Apply Air

Air calibration starts and will take 114 seconds to
complete. Under no circumstances interrupt the flow of
air through the flow channel.