Gas Flow Analyzer

Specifications

47

Specifications

Display ................................................................... 26 mm x 33 mm
Real time curves ................................................... Flow, Pressure, Volume, temperature of gas inside the measurement chamber, oxygen, respiratory

parameters

Interfaces ............................................................... RS-232, USB, Ethernet, CAN, Analog Out, TTL
Temperature (gas in measurement chamber)

Operating ............................................................ 15

°

C to 40

°

C (59

°

F to 104

°

F)

Storage ............................................................... -10

°

C to 60

°

C

Relative humidity

Operating ............................................................ 10 % to 90 % RH
Storage/Transportation ....................................... 5 % to 95 % RH

Ambient pressure ................................................. 500 mbar to 1150 mbar
Power

AC adapter

Voltage input ................................................... 100 V ac to 240 V ac, 50 Hz to 60 Hz
Supply voltage ................................................ 5 V dc
Power consumption ........................................ 2.5 W to 6 W

Battery

Battery life ....................................................... 4 hours. Operation time will be reached in standalone operation (without use of interfaces)
Recharge time ................................................ 5 to 8 hours (varies with port used)

Dimensions (W x L x H) ........................................ 16.5 cm x 10.8 cm x 6.4 cm (6.5 in x 4.25 in x2.5 in)
Weight .................................................................... 0.4 kg
Safety ..................................................................... IEC 61010-1: Pollution Degree 2
Electromagnetic Environment ............................. IEC 61326-1: Portable
Calibration interval ............................................... annually
Memory Card ......................................................... yes
Data Interfaces

Analog port

Analog Output 1 .............................................. 0 Vdc to 5 Vdc

±

1.8 %, load

5 k

Ω

Analog Output 2 .............................................. 0 Vdc to 5 Vdc

±

1.8 %, load

5 k

Ω