Connection to the respiratory apparatus, Standard trigger values, Baseflow – Fluke Biomedical VT305 User Manual

Page 53

Advertising
background image

Gas Flow Analyzer

Measure Respiratory Data

39

Connection to the Respiratory Apparatus

There are three different methods to connect the Product
to the respiratory apparatus:

Downstream of the Y-piece

Ventilator

Fluke VT305

Test Lung

gyo038.eps

In the inspiration duct upstream of the Y-piece

Ventilator

Fluke VT305

Test Lung

gyo039.eps

In the expiration duct upstream of the Y-piece

Ventilator

Fluke VT305

Test Lung

gyo040.eps

Standard Trigger Values

Because the Product can measure flow in each direction, it
makes sense to use the first connection method. In this
measurement setup, flow is usually chosen as the trigger
value. Flow triggers are stored as standard values in the
device and can be reset when necessary. The standard
trigger values for the flow trigger for adult breathing, for
example, are:

Starttrigger:

Flow > 3 l/min

Endtrigger:

Flow < -3 l/min

With the second and third connection methods, pressure is
usually chosen as the trigger signal. In this case the
standard values are as follows:

Starttrigger:

Pressure > 1 mbar

Endtrigger:

Pressure < 1 mbar

Baseflow

The base flow is the constant flow that must be ignored
when you calculate volume. If there is an identified
leakage in the system, like a constant discharge of 3 l/min