Fluke Biomedical VT305 User Manual

Page 62

Advertising
background image

VT305
Users Manual

48

Trigger Input ................................................... 5 Vdc to 24 Vdc
V

IN

.................................................................. 9 Vdc to 29 Vdc

RS-232 Port
Baud Rate ........................................................... 19200, 8 bits, no Parity, 1 stop bit

Measurement Variables

Air and N2

Flow Measurements

Range .........................................................

±

300 sl/min

Accuracy .....................................................

±

1.9 %* or

±

0.1 l/min

Ambient pressure compensated ..................... yes
Temperature compensated ............................ yes

O2/Air Mixtures

Flow Measurements

Range .........................................................

±

300 sl/min

Accuracy .....................................................

±

1.9 %* or

±

0.1 l/min

Ambient pressure compensated ..................... yes
Temperature compensated ............................ yes

CO2

Flow Measurements

Range .........................................................

±

140 sl/min

Accuracy .....................................................

±

3.0 %* or

±

0.1 l/min

Ambient pressure compensation accuracy ..... 25 °C to 30 °C
Temperature compensated ............................ yes
Channel pressure compensation accuracy .... -50 to +600 mbar

Heliox (21 % O2/ 79 % He)

Flow Measurements

Range .........................................................

±

300 sl/min

Accuracy .....................................................

±

4.0 %* or

±

0.3 l/min

Ambient pressure compensation accuracy ..... 25 °C to 30 °C
Temperature compensated ............................ yes