Notes -19 – Quantum Composers LDDC 1550 User Manual

Page 19

Advertising
background image

8-19 |

P a g e

:MR

1 - 100000

1 - 100000

Sets and/or queries the system's

Maximum Rate.

:OT?

0 = Okay

1 = Fault

Query only. Returns the Over

Temperature status.

:PE

0 or 1

0 = Disable
1 = Enable

Sets and/or queries the pulse mode

enable function.

:PM

0, 1, 2, 3

0 = CW
1 = Pulsed
2 = Burst
3 = Single

Sets and/or queries the system's Pulse

Mode.

:PW

200ns - 90%

of Rate

200ns - 90% of

Rate

Sets and/or queries the system's Pulse

Width.

:RC

1 - 5

1 - 5

Recall storage bin. Will recall the

settings in the specified user storage

bin.

:RR

0.1 - 100000

0.1 - 100000

Sets and/or queries the system's

Repetition Rate.

:SS?

16 bit decimal

Query Only. Returns the System's

State. See the System State section

below for return states.

:ST

0 or 1

0 = Disable
1 = Enable

Sets and/or queries the Start signal.

:SV

1 – 5

1 – 5

Save settings to specified storage bin.

:TB

0 or 1

0 = Disable
1 = Enable

Sets and/or queries the temperature

interlock bypass mode.

:VM?

0 – Compliance

Voltage

Query Only. Returns the Voltage

Measurement (Up to 3 decimal places).

:VN?

Ex: 1.2

Query Only. Returns the system
Version Number.

NOTES

(1)

The maximum Duty Cycle parameter allowed on the LDDC/LSC is a function of the Pulse

Mode and Period.

-In Burst Mode, the maximum allowable Duty Cycle percentage is:



-In Single Shot and Pulsed Mode, the maximum allowable Duty Cycle percentage is:

Advertising