Quantum Composers Jewel Laser User Manual

Page 24

Advertising
background image

Page 24 Jewel Manual v1.1 | Quantum Composers

SC?

System shot count, since factory build.

SS?

System Status – Query Only. Returns the current system state.

Value is a 16 bit decimal value with each bit position

corresponding to a system state (See Table below, bits 0-15).

15

14

13

12

11

10

9

8

Spare

Spare

Spare

Spare

Ready to

Fire

Ready

to

Enable

Power

Failure

Electrical

Over

Temp

(FET)

7

6

5

4

3

2

1

0

Resonator

Over

Temp

External

Interlock

Shutter

Closed

Q-Switch

External

Trigger

Diode

External

Trigger

Q-

Switch

Enabled

Laser

Active

Laser

Enabled

SV #

SV?

Save settings to user bin 1-6.

SY #

SY?

Sync Output Mode. 0 = off, 1 = LD pulse, 2 = QSW pulse, 3 =

LD/QSW.

TA?

Resonator thermistor A temp. In degrees C.

TB?

Resonator thermistor B temp. In degrees C.

UC #

UC?

User shot count. Saved shot count. Can be cleared on command

by issuing a 0 after command. Can also be preset by sending any

number up to 32 bit.

Harmonic Module Heater Commands (Note: Applicable if Installed)

H1?

Heater 1 installed (0 = No, 1 = Yes)

H1:SP?

Heater 1 setpoint. Target heater setting, in degrees C.

H1:ST?

Heater 1 Status byte (See Table Below).

H1:TP?

Heater 1 temperature. Active temperature of the heater in degrees

C.

H1:VN?

Heater 1 controller firmware version.

H2?

Heater 2 installed (0 = No, 1 = Yes)

H2:SP?

Heater 2 setpoint. Target heater setting, in degrees C.

H2:ST?

Heater 2 Status byte (See Table Below).

H2:TP?

Heater 2 temperature. Active temperature of the heater in degrees

C.

H2:VN?

Heater 2 controller firmware version.

7

6

5

4

3

2

1

0

spare

spare

spare

Temp

reading

out of

range

Temp

too high

At

Target

Temp

Temperature

settling

Temp

Control

Enabled