CAUTION: Use of controls, adjustments or performance of procedures other

than those specified in this Operation Manual may result in hazardous

radiation exposure.

Follow the instructions within this manual carefully to ensure the safe operation of your

laser. At all times during laser operation, maintenance, or servicing, avoid unnecessary

exposure to laser or collateral radiation that exceeds the accessible emission limits listed in

"Performance Standards for Laser Products," United States Code of Federal Regulations,

21 CFR 1040.10(d). This information is also available in EN60825-1:1994, Section 8.2,

titled “Measurements of Laser Radiation for Determining Classification.”

Preventative Maintenance for Safety: Preventative maintenance is required to ensure the

laser remains in compliance with Center for Devices and Radiological Health (CDRH)

Regulations and European Norm (EN) requirements. This laser product complies with Title

21 of the United States Code of Federal Regulations, Chapter 1, Subchapter J, Parts

1040.10, as applicable, and with EN60825-1:1994, Part 1 for a Class 4 laser, as applicable.

To maintain compliance, verify the operation of all features listed below, either annually or

whenever the product has been subjected to adverse environmental conditions, which may

have affected these features and functions.

o

Verify that removing the remote interlock connector prevents laser operation. This

connector is located on the end cap of the laser.

o

Verify that the laser will operate only with the key switch in the ON position and the

key can only be removed when the switch is in the OFF position.

o

Verify that a time delay exists between turning on the Key Switch and the start of

laser firing. The time delay must be long enough to allow action to be taken to avoid

exposure to laser radiation.

ELECTRICAL SAFETY

CAUTION: Both the Laser Optics Assembly and Laser Power Supply contain

electrical circuits operating at lethal voltage and current levels. Always

unplug the system Mains connection and wait at least one (1) minute to allow

capacitors to discharge before servicing any part of the laser system.

Consult with the Quantum Composers if repair of the laser electronics is required. Only those

trained in high voltage, high current electronics, and who understand the laser circuitry, should be

allowed to service and repair the laser electronics. If any such action is required, it is

recommended that you contact the Quantum Composers for details.