Watson-Marlow 520DuN IP66/NEMA 4X User Manual

32

Calibration dose

The pump offers to run a calibration dose. Use the UP and DOWN keys to
choose Yes or No. Press ENTER to confirm. Examples are shown here.
If No is selected after a tube selection has been made (see Tube above), the
pump briefly displays a confirmation screen showing current head, tube and
flow settings, and redisplays the main menu.
If Calibration dose is selected or if Yes is selected after a tube selection has
been made (see Tube above), the pump displays the speed and direction at
which it was last running in manual mode or to which it has just been set and
invites the user to press START.
Note: it is possible at this point to change the direction of rotation using the
DIRECTION key, and the change the speed of rotation using the UP and
DOWN keys, up to the pump’s maximum speed or any (lower) maximum avail-
able speed which has previously been set. See 18.13 Set maximum speed. If
the pumphead has been changed at this point, any previously set speed will be
over-ridden by any lower pumphead maximum speed.