3 sof tw a re set u p – HEIDENHAIN ND 1200R User Manual
Page 72

72
2 Installation and Specifications
2.3 Sof
tw
a
re
set
u
p
Position the standard artifact along the R axis.
Align the artifact as closely as possible to the axis and then perform
a skew alignment as described in chapter 1 (see "Aligning the part
to an axis" on page 27).
Perform a single measurement of the entire range of motion using
the standard artifact and make a note of the result.
In this example of applying LEC, one point at end of the axis
measurement range is measured using an 8 inch standard.
Use an artifact that allows measurements of as much of
the axis range of motion as possible.
LEC example using an 8 inch standard
0
0
+0.02
+0.01
0.00
-0.01
-0.02
8.0
7.98
1
2
3
4
Arrow number
Descriptions
1
: Standard length
Entire 8 inch length is measured
2
: Standard values
The certified length of the standard
3
: Observed values
The measured length of the standard
4
: Deviation graph
Difference between standard and observed
values (not entered into any screen)