Eagle Instruction Manual

Operation • 14

Setting User Access

The CAL/SETUP switch controls the Eagle functions available to the
user. The switch setting does not affect the Eagle’s ability to display
gas readings and indicate gas and malfunction alarms.

1. Turn off the Eagle.

2. Unscrew the two large screws on the top of the case.

3. Turn over the top half of the case, and locate the CAL/SETUP

switch (SW2) near the middle along one edge of the main
processor board.

CAUTION:

The Methane Elimination switch (SW1) is on the opposite edge
of the board near the front end. DO NOT confuse these two
switches.

4. Place the CAL/SETUP switch in the appropriate position.

• To give the Eagle access to all modes, place the switch in the

ON position.

• To limit the Eagle to normal operating and display modes,

place the switch in the OFF position. (The Eagle prevents
access to the setup and calibration modes by disabling the
SHIFT/▼ button.)

5. Place the top of the case in its original position, then secure it with

the large screws you loosened in step 2.

6. Turn on the Eagle.

NOTE:

Make sure the Eagle’s calibration is current and the setup
options are appropriate and safe for the operating
environment before placing the CAL/SETUP switch in the
OFF position.

Turning Off the Eagle

To turn off the Eagle, press and hold down the POWER/ENTER
button until

GOOD-BYE

displays, then release the button. (You must

wait for

GOOD-BYE

to disappear before you can turn on the Eagle

again.)