Eagle Instruction Manual
Maintenance • 52
5. Remove the sensor from the sensor block. If necessary, wiggle the
sensor to work it out of the seal.
6. Install the replacement sensor in reverse order.
CAUTION:
Verify that you install the H
2
S or CO sensor in the appropriate
flow block position. The Eagle will display inaccurate H
2
S and
CO readings if the sensors are not installed in the correct flow
block positions.
The Eagle continues to display H
2
S and CO readings in the
same channels as it did previously regardless of which socket
(EC1 or EC2) the sensor is wired to or into which flow block
position the sensor is installed. However, the H
2
S and CO
reading will be inaccurate due to the charcoal filter installed
before the CO flow block position.