RKI Instruments Eagle Series User Manual

Page 59

Advertising
background image

Eagle Instruction Manual

Maintenance • 52

5. Remove the sensor from the sensor block. If necessary, wiggle the

sensor to work it out of the seal.

6. Install the replacement sensor in reverse order.

CAUTION:

Verify that you install the H

2

S or CO sensor in the appropriate

flow block position. The Eagle will display inaccurate H

2

S and

CO readings if the sensors are not installed in the correct flow
block positions.

The Eagle continues to display H

2

S and CO readings in the

same channels as it did previously regardless of which socket
(EC1 or EC2) the sensor is wired to or into which flow block
position the sensor is installed. However, the H

2

S and CO

reading will be inaccurate due to the charcoal filter installed
before the CO flow block position.

Advertising