Hose and probe, Description – RKI Instruments Eagle Series User Manual

Page 65

Advertising
background image

Eagle Instruction Manual

Appendix C: Non-Standard Toxic Gas Sensors • 58

Hose and Probe

Some super toxic gases require a super toxic probe with a plastic
fitting instead of the standard hose with the metal fitting. This super
toxic probe does not have a particle filter installed. Use a short tubing
stub to connect the super toxic probe to the inlet of the Eagle. Other
super toxic gases cannot be measured with a probe at all. In both
cases, a 3 foot teflon hose is included with the instrument instead of
the standard hose. See the above table for the gases that fall into either
category.

Description

Non-standard toxic gas sensors are mounted in the front half of the
instrument case. These sensors each include a dedicated amplifier
with ZERO and SPAN controls. A bracket secures the sensor/
amplifier assembly to the instrument case. A flow adapter connected
to the sensor allows the sample to flow through the non-standard
toxic gas sensor. A seven-position connector plugs into the EC3
(channel 3) or EC4 (channel 4) socket on the analog PCB.

Hydrogen
Selenide**

0 to 200 ppb

1

50 ppb

100 ppb

50 ppb

100 ppb

Nitric Oxide

0 to 100 ppm

10 ppm

25 ppm

25 ppm

50 ppm

Nitrogen Dioxide

0 to 15.0 ppm

1.0 ppm

3.0 ppm

3.0 ppm

5.0 ppm

Ozone**

0 to 1.00 ppm

0.10 ppm

0.30 ppm

0.08 ppm

0.10 ppm

Phosphine

0 to 1.00 ppm

0.10 ppm

0.30 ppm

0.30 ppm

1.00 ppm

Silane

0 to 15.0 ppm

2.0 ppm

5.0 ppm

5.0 ppm

5.0 ppm

Sulfur Dioxide

0 to 10.0 ppm

1.0 ppm

2.0 ppm

2.0 ppm

5.0 ppm

Sulfur Dioxide

0 to 15 ppm

1.0 ppm

2.0 ppm

2.0 ppm

5.0 ppm

1

ppb (parts per billion)

*Use the super toxic probe (particle filter removed) when monitoring for these gases.

**Do not use a probe when monitoring for these gases. The probe will absorb the gas
sample.

Table 8: Non-Standard Toxic Gas Sensors Specifications

Target Gas

Range

Alarm 1

Alarm 2

TWA

Alarm

STEL

Alarm