37 • Setup Mode

Eagle Instruction Manual

2. Press and hold the SHIFT/▼ button, then press the DISP/ADJ

button. The Auto Calibration screen for the combustible gas
channel displays.

3. Press the AIR/▲ or SHIFT/▼ button to display the desired setting.

4. Press the POWER/ENTER button to enter the new setting. The

Auto Calibration screen for the next channel displays.

5. Repeat steps 4 and 5 for the remaining channels. (You must scroll

through all remaining channels before you can exit the Auto
Calibration screen.)

6. When the Auto Calibration screen for the last channel displays,

press the POWER/ENTER button to return to the main Auto
Calibration screen.

7. Press the POWER/ENTER button to return to the main menu.

Updating the Back Light Setting

This setting indicates the length of time the LCD illuminates when
you press any button. The minimum setting is off; the maximum
setting is 10 minutes. The default setting is 15 seconds.

1. From the main menu, select the LCD BACK LIGHT TIME menu

option.

2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired

setting.

3. Press the POWER/ENTER button to enter the setting and return to

the main menu.

Turning the Auto Fresh Air Function On or Off

WARNING: If Auto Fresh Air Adjust is ON, you must start the Eagle

in a “fresh-air” environment. If this setting is ON and
the Eagle is started in the presence of a target gas, the
readings and alarms will not be accurate or reliable.