Appendix d: carbon dioxide sensors, Specifications, Description – RKI Instruments Eagle Series User Manual

Page 71

Advertising
background image

Eagle Instruction Manual

Appendix D: Carbon Dioxide Sensors • 64

Appendix D: Carbon Dioxide Sensors

Appendix D describes the Eagle’s non-standard, infrared carbon
dioxide (CO

2

) sensors. It also offers additional information for

calibrating and replacing CO

2

sensors.

Specifications

Table 10 lists specifications for the carbon dioxide sensors. The alarm
settings are user-adjustable (see “Updating the Alarm Point Settings”
on page 33).

Description

The Eagle uses an infrared sensor to detect carbon dioxide. The carbon
dioxide sensor includes a dedicated amplifier with ZERO and SPAN
controls. A bracket secures the sensor/amplifier assembly to the
instrument case. A four-position connector plugs into the EC3
(channel 3) or EC4 (channel 4) socket on the analog PCB.

A carbon dioxide scrubber is mounted to the exterior front of Eagles
that are factory-shipped with carbon dioxide sensors for the ranges 0 -
5,000 ppm, 0 - 10,000 ppm, and 0 - 5%.

NOTE:

Eagles with a range of 0 - 20% and 0 - 60% CO

2

do not include

a scrubber since the normal background of CO

2

in air is

negligible when compared to the full scale of these units.

Table 10: Carbon Dioxide Sensor Specifications

Range

Alarm 1

Alarm 2

TWA Alarm

STEL Alarm

0 - 5000 ppm

OFF

OFF

OFF

OFF

0 - 10,000 ppm

5000 ppm

OFF

5000 ppm

OFF

0 - 5.00% CO2

0.50% CO2

3.00% CO2

0.50% CO2

3.00% CO2

0 - 20.0%

OFF

OFF

OFF

OFF

0 - 60.0%

OFF

OFF

OFF

OFF

Advertising