RKI Instruments Eagle Series User Manual

Page 38

Advertising
background image

31 • Setup Mode

Eagle Instruction Manual

5. Press the AIR/▲ or SHIFT/▼ buttons to display the available

target gas labels for the toxic gas channel (H2S, CO, SO2, Cl2,
NH3

, CO2 (5.00%), CO2 (10000 PPM), CO2 (5000 PPM), ***, and

NOT USED

).

6. Press the POWER/ENTER button to enter the new target gas label.

If you entered a label other than ***, continue with step 7. If you
entered ***, go to the next section, “Setting a custom target gas label.”

7. To exit the Gas Combinations menu, press the SHIFT/▼ button

until the prompt is next to Channel 4, then press the SHIFT/▼
button again. The

ESCAPE

message displays.

8. Press the POWER/ENTER button. The message

SAVING DATA

displays, then the main menu displays.

Setting a Custom Target Gas Label

1. With the prompt next to the target gas label setting (***), press the

POWER/ENTER button. The first asterisk flashes.

2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired

character. Available characters are A through Z, 0 through 9, and a
blank space.

3. Press the POWER/ENTER button to enter the displayed character.

The next character flashes.

4. Repeat steps 2 and 3 to enter the remaining characters. When you

enter the last character, the prompt flashes.

Updating the Fullscale Setting

1. Press the SHIFT/▼ button to place the prompt in the second line,

then press the POWER/ENTER button. The fullscale setting
flashes.

The maximum fullscale setting for a toxic gas channel is 1000 PPM;
the minimum setting is 1.00 PPM. The default setting is 10.0 PPM.

2. Press the AIR/▲ and SHIFT/▼ buttons to display the desired

fullscale setting, then press the POWER/ENTER button to enter
the setting. The prompt flashes.

NOTE:

The display increment setting automatically updates its
default setting as you change the fullscale setting.

* * *

1

.

P

0

0

>

0

P M

1

.

P P M